AMO Tecnis 1-Piece Intraocular Lens (Model ZCB00)

Company
Abbott Medical Optics Inc (Amo)
Recall Initiated
November 11, 2009
Posted
January 12, 2010
Terminated
July 25, 2012
Recall Number
Z-0605-2010
Quantity
246000 Lenses
Firm Location
Santa Ana, CA

Reason for Recall

Abbott Medical Optics (AMO) has received complaints of some Tecnis 1-Piece IOLs (model ZCB00) have adhered to the lens optic after insertion into the eye. This can lead to a need for additional manipulation to separate the haptics from the optic during surgery which could lead to an injury.

Distribution

AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, Hawaii, and Puerto Rico. INTERNATIONAL: Austria, Azerbaijan, Belgium, Canada, Switzerland, Cyprus, Czech Republic, Germany, Estonia, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Italy, Lebanon, Lichtenstein, Libya, Martinique, Netherlands, Poland, Palestine, Portugal, Saudi Arabic, Sweden, Slovenia, Turkey, and South Africa.

Lot / Code Info

All Serial numbers of tecnis 1-Piece Intraocular Lenses that were processed through lens surface treatment at the AMO PR facility between May 19 and October 16, 2009 - with serial numbers ending in 0906, 0907, 0908, and 0909. and subset with serial numbers ending in 0905 and 0910.  Expiration/Use By Date (YYYY-MM): 2011-05, 2011-06, 2011-07, 2011-08, 2011-09, and 2011-10

Root Cause

Process design

Action Taken

Beginning on Friday, November 6, 2009, AMO began distribution of the Recall Notification Letter via Federal Express to the 874 AMO Tecnis 1-piece IOL customer accounts in the US. Included with the Recall Notification letter is a facsimile delivery confirmation that each customer was instructed to fax back to AMO to confirm receipt of the Recall Notification Letter. --- Beginning on Monday, November 9, 2009, AMO began distribution of the Recall Notification letter to the AMO Tecnis 1-piece IOL customer/distributors accounts outside the USA. "THIS RECALL NOTIFICATION DOES NOT INVOLVE TECNIS 1-PIECE IOL THAT HAVE BEEN IMPLANTED. ONCE THE HAPTICS HAVE BEEN SUCCESSFULLY DEPLOYED, THESE LENSES FUNCTION PROPERLY." Questions should be directed to an AMO Customer Service Representative at 1-877-AMO-4LIFE (1-877-266-4543).