Various Types and Sizes of Polyethylene Synthetic Nonabsorable Sutures-Polyester, TEV GN BR 0 TC-43/HR 262N, Catalog #833-114

This recall has been terminated (originally issued July 17, 2007).

Recall Initiated
June 6, 2007
Posted
July 17, 2007
Terminated
March 14, 2008
Recall Number
Z-1017-2007
Quantity
1920 units total for all types of sutures
Firm Location
Durham, NC
Official Source
View on FDA website ↗

Reason for Recall

Suture Diameter/Strength specification failure: The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage.

Distribution

Worldwide-, USA including states of CO, FL, IL, IN, MA, and OH, and Countries of Canada, France and Germany

Lot / Code Info

Polyester, TEV GN BR 0 TC-43/HR 262N, Catalog #833-114, Lot #21486, 21487, 21488, 23468, 23469, 41910 & 55380.

Root Cause

Other

Action Taken

Consignees were notified by letter on/about 06/07/2007. They were advised to cease use, remove and quarantine all affected product immediately. Affected products are to be returned to Teleflex Medical. An Acknowledgement & Stock form was included to be completed and returned via fax to Teleflex Medical.