C1140 Flexible Endoscope Processing Container and Tray assembly.
- Company
- Steris Corp
- Recall Initiated
- April 23, 2003
- Posted
- May 15, 2003
- Terminated
- February 5, 2004
- Recall Number
- Z-0830-03
- Quantity
- 354 units
- Firm Location
- Mentor, OH
Reason for Recall
The container flow port is inverted, which restricts the flow of sterilant to the endoscope and affects the sterilization process.
Distribution
The product was shipped to end-users located nationwide and worldwide.
Lot / Code Info
Serial Numbers: 203793 - 204362
Root Cause
Other
Action Taken
The firm notified their customers by telephone and letter on 4/23/2003.
More recalls by Steris Corp
STERIS Harmony LL 500/700 Surgical Light, Vertical Spring-Loaded Arm.
Intended for local illuminati...
Nov 23, 2009
Cmax 110v Surgical Table Hand Control.
The Cmax Surgical Table is a mobile, electro-hydraulically o...
Jul 21, 2009
ASC2000 Ambulatory Surgical Table. Model #MTS110003.
The device is used to provide complete and f...
Jul 7, 2009