ORTHOFIX Centurion POCT System Drill Guide PN 69-1010 and PN 69-1012. Drill Guides are used to interface with various drill tips during the drilling procedure of bone preparation.

This recall has been terminated (originally issued June 19, 2015).

Company
Orthofix, Inc
Recall Initiated
May 11, 2015
Posted
June 19, 2015
Terminated
February 17, 2016
Recall Number
Z-1819-2015
Quantity
36 units
Firm Location
Lewisville, TX
Official Source
View on FDA website ↗

Reason for Recall

There is a possibility that the thumb button remains in the depressed (unlocked) position allowing the threaded sleeve (drill stop) to unintentionally rotate and change the depth setting.

Distribution

Nationwide Distribution including CA, VA, FL, ID, CO, OR, PA, OK, SC, and TX.

Lot / Code Info

All lot codes

Root Cause

Device Design

Action Taken

Orthofix sent an Urgent Medical Device Product Advisory Notification letter dated May 11, 2015, to the affected distributors and hospitals. The consignees were directed to inspect thier inventory and perform checks before surgical use of the drill guides. Customers were instructed to provide a copy of the Product Advisory Notification letter if product was further distributed. Customers with questions were instructed to contact their local Orthofix representative and to for returns they should contact Orthofix customer service representative at 888-298-5700.