RecallDepth

ACCU-CHEK Connect Diabetes Management App; Instruction Manual Designed to transfer data for diabetes management.

Company
Roche Diabetes Care, Inc.
Recall Initiated
October 30, 2014
Posted
April 2, 2015
Terminated
June 8, 2015
Recall Number
Z-1369-2015
Quantity
The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.
Firm Location
Indianapolis, IN

Reason for Recall

Roche Diabetes Care has become aware of an issue with the Accu-Chek Connect diabetes management app that could potentially lead to inaccurate bolus advice being provided to the user. A thorough investigation of the situation revealed that this issue may occur if the user changes the screen orientation of the phone from portrait to landscape or vise versa while looking at the Bolus Advisor or Carb

Distribution

There are no US distributors as the affected device is not marketed in the US. The affected device was distributed in Germany, Italy and South Africa. The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.

Root Cause

Software design

Action Taken

Urgent Field Safety Notices were sent to the customer and healthcare professional starting on 10/30/2014. The letter provided a description of the reason for the notification; and recommended that all users should update their app immediately.

More recalls by Roche Diabetes Care, Inc.

View all recalls by this company →