Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG) Hemostasis System.

This recall has been terminated (originally issued February 19, 2015).

Company
Haemonetics Corporation
Recall Initiated
January 26, 2015
Posted
February 19, 2015
Terminated
June 5, 2015
Recall Number
Z-1130-2015
Quantity
4,143 kits
Firm Location
Braintree, MA
Official Source
View on FDA website ↗

Reason for Recall

Specific lots of Level II Control Kits produce high Maximum Amplitude (MA) results, which are outside of the manufacturer's expected Quality Control range for MA parameter.

Distribution

Worldwide Distribution -- US, Canada, Czech Republic, Denmark, Great Britain, and Sweden.

Lot / Code Info

Lot numbers: HMO 9200  HMO 9217  HMO 9226  HMO 9232  HMO 9276  HMO 9285  HMO 9292 Exp. Date 1/2017

Root Cause

Process control

Action Taken

Haemonetics issued on 1/26/15, an Urgent Medical Device Recall to consignees The letter identified the affected product, the reason for the recall, as well as the risk to health. It instructs consignees to inspect their stock and determine if they have any of the affected lots. Customers are to immediately discontinue use and discard as instructed any affected product. The attached customer acknowledgment form should be completed and returned whether or not customers have any affected product. Customers are to contact their local service representative at 1-800-537-2802 if replacement product is needed. If any affected product has been forwarded to other locations, customers are to inform them of the recall. Please call the Technical Support team at 1-800-438-2834, with questions about this action.