RecallDepth

GuardIVa Antimicrobial Hemostatic IV Dressing, REF FP-23-AD008, Manufacturer: Bard Access Systems, Inc. USA. Product Code FP23AD006.

This recall has been terminated (originally issued December 1, 2014).

Company
Bard Access Systems
Recall Initiated
November 3, 2014
Posted
December 1, 2014
Terminated
May 12, 2015
Recall Number
Z-0493-2015
Quantity
2532 units
Firm Location
West Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

Bard Access Systems is recalling GuardIVa (Ref. No. FP23AD006) from lot ASXHT003 because they were non-sterile unit samples which potentially could be used clinically.

Distribution

US Distribution to state of: TX, OK, LA, and AR.

Lot / Code Info

Product Code FP23AD006, Lot No.ASXHT003, Expiration Date Feb 2016

Root Cause

Mixed-up of materials/components

Action Taken

A customer letter was sent on 11/3/14 requesting immediate examination of inventory and to quarantine the products for return.

More recalls by Bard Access Systems

View all recalls by this company →