Aquatec Fixed Backrest - accessory for Mobile Shower and Toilet commode Aquatec Ocean, Ocean VIP and Ocean E-VIP family.
- Company
- Invacare Corporation
- Recall Initiated
- June 25, 2014
- Posted
- July 9, 2014
- Terminated
- April 15, 2015
- Recall Number
- Z-2023-2014
- Quantity
- 465 units
- Firm Location
- Elyria, OH
Reason for Recall
The Aquatec Fixed Backrest withstands a load of only 250 newtons at uneven load. This may result in the Fixed Backrest becoming detached from the frame.
Distribution
Distributed in PA and Canada.
Lot / Code Info
Mobile Shower and Toilet commode Aquatec Ocean (Model # 1534327), Ocean VIP (Model # 1470713 and 1470793) and Ocean E-VIP (Model # 1525707).
Root Cause
Device Design
Action Taken
The recall notification will be sent to the two consignees informing them of the recall and to notify their customers. The letter will include a provider letter, customer letter, provider response card and impacted units listing.