Vac-Assist Suction Units (SUA01 Suction Aspirator Unit w/800 cc Canister). The suction unit is intended for professional use to remove fluids from a patient's airway or respiratory support unit. The intended population for this device is s both adult and pediatric.

This recall has been terminated (originally issued September 13, 2014).

Company
Emg Technology Co., Ltd.
Recall Initiated
August 3, 2014
Posted
September 13, 2014
Terminated
August 25, 2015
Recall Number
Z-2642-2014
Quantity
1452
Official Source
View on FDA website ↗

Reason for Recall

suction canister lids may have occluded ports.

Distribution

US Distribution in IL.

Lot / Code Info

Product Model No: SUAOI Serial Number ranges: SUI4B01391-SU14B02117. SU14C00556-SU14C01280

Root Cause

Nonconforming Material/Component

Action Taken

A recall letter, dated August 6, 2014, was sent via email which identified the product, problem, and actions to be taken.