Heparin I.V. Flush Syringe is available is pre-filled polypropylene syringes, 100 units/mL, 3 mL fill in a 12 mL syringe.

This recall has been terminated (originally issued March 6, 2014).

Company
Medefil Incorporated
Recall Initiated
February 19, 2014
Posted
March 6, 2014
Terminated
September 5, 2017
Recall Number
Z-1149-2014
Quantity
1,608 Master Cartons
Firm Location
Glendale Heights, IL
Official Source
View on FDA website ↗

Reason for Recall

The bar code that is printed on the Master Carton label does not correspond to the product code fully displayed on the Master Carton label. The non-matching information can only be detected by using a bar code reader. The defective bar codes will be read as Product No. MIH-3335 ( Heparin I.V. Flush Syringe 100 units/mL; 5 mL fill in 12 mL Syringe) whereas this product is actually Product No. MIH-3333 (Heparin I.V. Flush Syringe 100 units/mL; 3 mL fill in 12 mL Syringe).

Distribution

USA (nationwide) including the states of AZ, CA, CO, FL, GA, IL, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, SC, TX, UT, and WI, and the country of Canada.

Lot / Code Info

Product Number: MIH-3333;  NDC Number: 64253-333-33;  Lot Numbers and Expiration Dates: Lot H114120N, Expires 12/15; Lot H113403N, Expires 11/15; Lot H113369N, Expires 10/15; Lot H113327N, Expires 8/15; Lot H113291N, Expires 7/15; Lot H113266N, Expires 6/15; Lot H113232N, Expires 5/15; Lot H113185N, Expires 3/15.

Root Cause

Labeling Change Control

Action Taken

URGENT MEDICAL DEVICE CORRECTION letters dated February 19, 2014 were sent to all direct accounts. The letters included instructions to: quarantine the recalled products; and,complete and return the attached customer response form (via Fax to 630-681-9100 or via mail to Medefil Inc., 250 Windy Point Drive, Glendale Heights, IL 60139) so that the correct number of replacement labels can be sent from Medefil.