VERIFY Dual Species Self-Contained Biological Indicator, Distributed by STERIS Corporation, Mentor, OH. Intended for use in installation testing and routine monitoring of steam sterilization and ethylene oxide sterilization processes.

This recall has been terminated (originally issued December 28, 2013).

Company
Steris Corporation
Recall Initiated
September 10, 2013
Posted
December 28, 2013
Terminated
February 14, 2014
Recall Number
Z-0605-2014
Quantity
1,238 boxes
Firm Location
Mentor, OH
Official Source
View on FDA website ↗

Reason for Recall

STERIS has identified that the population of the G. stearothermophilus spores that monitor STEAM sterilization cycles does not meet specifications for Lot # 131004 of the Verify Dual Species Self-Contained Biological Indicators. As a result the spore count on the affected product is lower than the labeled minimum and the product may incorrectly indicate that a 5-log spore kill was achieved.

Distribution

Worldwide Distribution-USA (nationwide) including the states of AK, Al, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY, and the countries of BAHRAIN, CANADA, COLUMBIA, GERMANY, ITALY, MALAYSIA, PHILIPPINES, ROMANIA, SPAIN, and THE NETHERLANDS.

Lot / Code Info

Lot #131004 Model #'s: S3060, S3061, S3065, S3069, LCB006 & LCB007.

Root Cause

Use error

Action Taken

On September 10, 2013, the firm sent Urgent Voluntary Recall Notices to their customers.