Draeger Movita 603x/DVE 803x Ceiling Supply Unit. It is intended for use in operating theaters and intensive care

This recall has been terminated (originally issued November 14, 2013).

Company
Draeger Medical, Inc.
Recall Initiated
September 30, 2013
Posted
November 14, 2013
Terminated
September 3, 2014
Recall Number
Z-0282-2014
Quantity
139
Firm Location
Telford, PA
Official Source
View on FDA website ↗

Reason for Recall

Cases were encountered with the Draeger Movita 603x/DVE 803x Ceiling Supply Units where the drive screw in the motor unit broke and the lift arm fell down approximately 60 cm (23.6 inches).

Distribution

Worldwide distribution: US (Nationwide)and in the country of Canada. (note: the ceiling supply units are not classified as a medical device by Health Canada).

Lot / Code Info

Catalog numbers:MZ00614, MZ01038, MZ02190, MZ02254, and MZ02332

Root Cause

Equipment maintenance

Action Taken

An "Urgent Medical Device Recall" notification letter was sent to users in September 2013. The issue was described and recommended actions were provided. Customers may contact (800) 543-5047 and select 1, then 2.