RecallDepth

Ultra-Thin LightMat Surgical Illuminator, Disposable one time use. Model Number: UA2550, Exp Date: 2012-05. Product is packaged in individual sterile pouch w/5 pouches per box. The LightMat is intended for the illumination of surgical procedures, particularly where deep cavities or adjacent tissues limit outside light in the surgical field.

This recall has been terminated (originally issued September 3, 2009).

Company
Lumitex Inc
Recall Initiated
July 1, 2009
Posted
September 3, 2009
Terminated
March 9, 2012
Recall Number
Z-1998-2009
Quantity
59 packs/5 individual units
Firm Location
Strongsville, OH
Official Source
View on FDA website ↗

Reason for Recall

Adhesive which is used to attach LightMat to retractor may not have been tested to Medical standards. The vendor provided an adhesive strip that did not conform to Lumitex specifications. The adhesive strip provided may not have met Lumitex requirements to be medical grade, EtO compatible, and biocompatible.

Distribution

Product was distributed to the following states: CA, FL, KS, MA, ME, MI , NC, NM, NV, NY, OH, and TX.

Lot / Code Info

Model Number: UA2550; Lot Number: I01161.

Root Cause

Nonconforming Material/Component

Action Taken

Lumitex, Inc. issued a recall letter dated July 2, 2009 to their Customers describing the affected device and requesting its return. For further information, contact Lumitex, Inc. at 1-440-243-8401.

More recalls by Lumitex Inc

View all recalls by this company →