iPulse Circulatory Support System Catalog number: 0036-0010 The iPulse System Console is a multi-purpose mechanical circulatory support system used to support patients suffering from acute cardiovascular failure. The iPulse console can be used to drive exclusively one of two classes of therapeutic devices: either pneumatically actuated Ventricular Assist Devices (VADs) for total circulatory support, or (i.e., non-simultaneously) an intra-aortic balloon (IAB) catheter for counter-pulsation therapy.

This recall has been terminated (originally issued September 16, 2009).

Company
Abiomed, Inc.
Recall Initiated
April 23, 2009
Posted
September 16, 2009
Terminated
June 5, 2014
Recall Number
Z-1931-2009
Quantity
79
Firm Location
Danvers, MA
Official Source
View on FDA website ↗

Reason for Recall

iPulse Console SC1035 displayed a "Low Pressure / Low Flow" alarm and stopped pumping during patient transport in battery operation.

Distribution

Domestic: MA, MI, SC, NY, NJ, OH, CA, PA, AR, MI, MD, CT, IL, WI, TX, NJ, TN, IN, MI, DE, FL, IA, WA, KY, and WV Foreign: Germany, Greece, Italy, Latvia, Spain, and Turkey

Lot / Code Info

Serial numbers: SC1000, SC1001, SC1003 - SC1010, SC1014 - SC1023, SC1025 - SC1070, SC1072 - SC1081, SC1083, SC1084, and SC1089

Root Cause

Device Design

Action Taken

An "Urgent Voluntary Medical Device Field Correction Notice" was issued April 8, 2009 to affected customers. The customers were notified that the iPulse Console SC1032 displayed a "Low Pressure/ Low Flow" alarm and stopped pumping during patient transport in battery operation. Instructions were provided to users in case they encounter this failure mode. ABIOMED Field Service Engineers will contact customers to schedule a time to visit site and correct affected consoles. Please contact ABIOMED's Field Service at 1.800.554.8666, if you have any questions.