RecallDepth

Surgidyne S-Vac round fluted channel silicone drain, 1/8" trocar, 10 FR, round, full fluted, 28 cm length, 4 channel, contains 10, sterile, Sterion Inc., Ham Lake, MN; REF 332187. The device is a sterile, portable closed system designed to evacuate post-operative wound drainage and promote healing.

This recall has been terminated (originally issued May 22, 2009).

Company
Aspen Surgical Products, Inc.
Recall Initiated
April 13, 2009
Posted
May 22, 2009
Terminated
November 23, 2009
Recall Number
Z-1638-2009
Quantity
9 cases of both products
Firm Location
Caledonia, MI
Official Source
View on FDA website ↗

Reason for Recall

An adaptor, which may be required when connecting the drain to an external device, was not included in the package.

Distribution

Nationwide Distribution -- California, Florida, Georgia, Massachusetts and Missouri.

Lot / Code Info

Lot Number 2014-02-13.

Root Cause

Process change control

Action Taken

An "Urgent Medical Device recall" letter dated April 13, 2009 was issued to consignees via certified mail. The letter described the affected product. and actions for customers. Consignees were instructed to carry out the recall to the retail level and to quarantine and return the product to Aspen Surgical Products, Inc. Consignees were also instructed to complete the enclosed response form and return it via mail, fax or e-mail. Direct questions about the recall by calling Aspen Surgical Products, Inc. at 1-616-698-7100, extension 142 or extension 141.

More recalls by Aspen Surgical Products, Inc.

View all recalls by this company →