Boston Scientific brand Position Acquisition Module (PAM), a component of Realtime Position Management (RPM) System, Catalog/Model Numbers: 8200010, 8200TWA0

This recall has been terminated (originally issued October 19, 2005).

Company
Boston Scientific
Recall Initiated
February 18, 2005
Posted
October 19, 2005
Terminated
January 30, 2006
Recall Number
Z-0064-06
Firm Location
San Jose, CA
Official Source
View on FDA website ↗

Reason for Recall

The device has the potential for overheating its transformer and has the potential for fire hazard.

Distribution

Product was released for distribution to 38 consignees in US nationwide. International distribution includes EU and Canada. It did not involve Defense Supply Center, VA or other Federal Government sales/distribution centers.

Lot / Code Info

All lots

Root Cause

Other

Action Taken

The firm sent notification letters to its consignees on 2/18/05 and 5/18/05 via FedEx. On the letter, a service reply form is included to monitor responses and schedule appointments to replace the PAM unit component.