RecallDepth

BIPLANE X-ray Imaging Systems included Allura XPER FD10/10, FD20/10, FD20/20, Integris BH5000, and Integris Allura 12-15 Biplane. The BIPLANE system has two C-arms and it allows the operator to image in two planes: Lateral and Frontal. It has a C-arm on the stand and C-arm suspended from the ceiling. The product's brochure is labeled in part: "Philips Healthcare Global Information Center...P.O. Box 1286...5602 BG Eindhoven...The Netherlands".

This recall has been terminated (originally issued September 22, 2010).

Company
Philips Medical Systems North America Co. Phillips
Recall Initiated
February 4, 2010
Posted
September 22, 2010
Terminated
September 23, 2011
Recall Number
Z-2449-2010
Quantity
267
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted.

Distribution

Nationwide Distribution

Lot / Code Info

Each unit is identified with a SITE number:  76992, 103526, 103581, 103921, 104020, 104021, 104068, 104386,  104497, 104628, 104926, 104950, 104953, 104959, 105184, 105399, 105581, 105585, 105654, 504655, 504674, 504733, 504742, 504779, 504868, 504921, 505067, 505068, 505120, 505356, 505380, 505530, 505611, 505703, 505903, 506008, 506026, 506069, 506119, 506509, 506655, 519118, 519136, 519196, 519197, 519678, 519730, 519981, 520275, 520582, 520666, 520688, 520756, 520793, 520949, 520964, 521149, 521325, 521525, 522136, 522456, 530396, 530653, 530654, 530660, 531649, 531758, 531799, 531926, 531939, 532152, 532218, 532240, 532241, 532383, 532614, 533288, 533290, 533397, 534131, 534156, 534786, 535345, 535761, 535775, 536028, 536031, 536034, 536035, 536037, 536042, 536171, 536715, 536789, 536899, 537175, 537593, 537631, 537804, 538028, 538095, 538099, 538157, 538172, 538387, 538409, 538514, 538516, 539213, 539326, 539628, 540184, 540237, 540461, 540721, 541237, 541238, 541239, 541276, 541577, 541589, 541590, 542382, 542413, 542557, 542714, 542963, 543289, 543411, 543690, 543879, 543953, 543989, 544306, 544395, 544543, 544635, 544656, 544668, 544844, 545025, 545262, 545396, 545518, 545583, 545604, 545753, 545882, 545884, 545997, 546013, 546119, 546149, 546569, 546906, 547173, 547215, 547431, 547642, 547715, 547769, 547823, 547922, 548005, 548020, 548036, 548082, 548140, 548288, 548383, 548618, 549202, 549242, 549368, 549370, 549903, 550198, 550691, 550758, 550812, 550935, 551855, 552355, 552599, 552639, 552969, 553685, 553699, 553766, 554763, 554802, 554970, 555293, 555299, 555356, 555834, 558604, 558945,  41443586, 41443609, 41443627, 41443647, 41443727, 41443765, 41443804, 41443909, 41444645, 41444647, 41444650, 41444689, 41444752, 41444816, 41445118, 41445432, 41445448, 41445501, 41445538, 41445615, 41445628, 41446230, 41446262, 41446773, 41455885, 41455897, 41455920, 41455925, 41455933, 41455944, 41455953, 41660520, 41780160, 41793290, 41795344, 41796154, 42046093, 42046440, 42057324, 42064058, 42067824, 42394072, 42479274, 42508759, 42553772, 42564341, 42625859, 42735090, 42891020, 42891095, 42891704, 43043666, 43058934, 43164789, 43173591, 43315510, 43436747, 43443622, 43453735, 43635004, 43670742, 43848923, 44028554, 44066424, 44584972, 44616211, 44654768, 44735751, and 44892370.

Root Cause

Software design

Action Taken

On 02/04/2010, Philips Healthcare began sending out the URGENT-Device Correction Notices to their consignees. The firm issued two notifications, one for the Allura Integris and Xper FD10C, FD10F, FD10F/C and FD20 Monoplane X-Ray imaging systems with the Velara Generator. The second notification is for the Allura Integris Xper FD10 and FD20 Biplane X-ray imaging systems with the Velara Generator. This notice informs the consignees that a system locks up during a critical interventional case and it might impose a serious injury to patients. This lock up issue could only be resolved by a warm or cold system restart. A warm restart would take about 90 seconds or a cold restart would take about four minutes, and the device could not be used during this time. If the system does not work after the restart, the consignees should contact the Philips Healthcare Customer Service Representative at 1-800-772-9377 #5 (then follow prompts). A service representative will visit each consignee and conduct a software and hardware upgrade.