RecallDepth

Byrd Screw, 2.0X6MM SD, Model Number: 01-0264 Product labeling reads in part as follows: REF: 01-0264, QTY,UOM 00001, BYRD SCREW, 2 OX6MM SELF-DRILLING, 12CM, (4-3/4) (L), STAINLESS STEEL, LOT 999999, (01) 00841036007157(10)999999(13)090201(30)13, 2009-02, ***NON-STERILE PRODUCT***DD 14 03 1, BIOMET MICROFIXATION, INC., 150 TRADEPORT DRIVE, JACKSONVILLE, FL 32218 USA Instrument used to pull up the broken bone, so it is flush and can be plated fixated. The Byrd screw is then removed.

This recall has been terminated (originally issued November 10, 2009).

Company
Biomet Microfixation, Inc.
Recall Initiated
February 19, 2009
Posted
November 10, 2009
Terminated
May 3, 2012
Recall Number
Z-0169-2010
Quantity
22
Firm Location
Jacksonville, FL
Official Source
View on FDA website ↗

Reason for Recall

Byrd Screw, Self Drilling, may break during usage. BIOMET has determined that the parts were not correctly processed by the vendor , i.e. that the parts were not hardened in the final processing and that lead to weaker than acceptable material.

Distribution

AZ, CA, GA, IL, IN, NM, NY, TX, VA, Puerto Rico and Canada.

Lot / Code Info

Lot # 091807107

Root Cause

Nonconforming Material/Component

Action Taken

Biomet plans to send letters to customers on February 19, 2009. Customers are told to discontinue use of the item and return it to Biomet Microfixation. The firm plans to follow up with an e-mail to their customers. Questions pertaining to this issue, should be directed to Christy Cain at 1.800.874.7711 or 904.741.4400.

More recalls by Biomet Microfixation, Inc.

View all recalls by this company →