Warfarin Metabolism Nucleic Acid Test Cartridge; Part Number: 20-006-002; an in vitro diagnostic drug metabolizing enzyme genotyping system; Nanosphere, Inc., 4088 Commercial Ave., Northbrook, IL 60062

This recall has been terminated (originally issued March 31, 2008).

Company
Nanosphere, Inc.
Recall Initiated
February 25, 2008
Posted
March 31, 2008
Terminated
May 17, 2010
Recall Number
Z-1383-2008
Quantity
624 units
Firm Location
Northbrook, IL
Official Source
View on FDA website ↗

Reason for Recall

Aberrant Results: A warfarin 2C9*2 mutant capture signal was found to be high enough to result in an aberrant result (i.e., wild-type called heterozygous).

Distribution

Nationwide; USA including states of Kentucky, Michigan, Tennessee, Texas, Louisiana, Connecticut, Illinois, and California.

Lot / Code Info

Lot Numbers 121907002B, 010708002A, 012108002A, 012108002C, 012208002C, 012408002B, 012808002A, and 020608002A

Root Cause

Other

Action Taken

Nanosphere sent Urgent Device Recall letters dated 2/25/08 to the direct accounts on the same date, informing them of the potential for aberrant results with the use of the affected lots of cartridges. The accounts were requested to examine their stocks for the affected lots of cartridges, segregate those found and return them to Nanosphere for replacement. Any questions can be directed to Dr. Gregg Shipp at 847-400-9115.