RecallDepth

Symbia S Series SPECT (Single-Photon Emission Computed Tomography) System with a Pinhole Collimator; Siemens Medical Solutions USA, Inc., Molecular Imaging, 2501 N. Barrington Road, Hoffman Estates, IL 60195-5203 USA; part number 08717741

This recall has been terminated (originally issued February 23, 2007).

Company
Siemens Medical Solutions Usa, Inc.
Recall Initiated
January 30, 2007
Posted
February 23, 2007
Terminated
November 30, 2007
Recall Number
Z-0548-2007
Quantity
16 units
Firm Location
Hoffman Estates, IL
Official Source
View on FDA website ↗

Reason for Recall

Symbia systems, running on e.soft version 5.5 or earlier, could allow a fully-extended patient pallet to come into contact with the pinhole collimator, resulting in system damage.

Distribution

Worldwide distribution --- including USA states of Alabama, California, Georgia, Indiana, Kentucky, Michigan, Ohio, Pennsylvania, Texas, Virginia and countries of Australia, Bolivia, Canada, China, France, Germany, Hong Kong and Spain.

Lot / Code Info

All systems with pinhole collimators and e.soft software versions 5.5 or earlier: part number 08717741, serial numbers 01006, 01008, 01009, 01019, 01021, 01022, 01024, 01032, 01033, 01034, 01038, 01044, 01045, 01049, 01051, 01052

Root Cause

Other

Action Taken

Siemens sent Customer Safety Advisory Notice/Urgent Field Correction Recall letters dated 1/30/07 to all users of Symbia systems with a pinhole collimator and e.soft software versions 5.5 or earlier. The accounts were informed of the potential for a fully-extended patient pallet to come into contact with the pinhole collimator, resulting in system damage. The customers were advised that a Siemens Customer Service Representative would be sent to their site to upgrade their system software to version syngeo MI Applications 2006A within 30 days. The upgrade will be done free of charge.

More recalls by Siemens Medical Solutions Usa, Inc.

View all recalls by this company →