Hamilton brand RAPHAEL Ventilator (Software Versio 2.2x), Product is distributed by Hamilton Medical, Inc., 4990 Energy Way,Reno, NV 89502-Serial Numbers 2975 to 5360,

This recall has been terminated (originally issued July 22, 2006).

Company
Hamilton Medical Inc
Recall Initiated
May 24, 2006
Posted
July 22, 2006
Terminated
December 21, 2006
Recall Number
Z-1249-06
Quantity
2,385 for all models
Firm Location
Reno, NV
Official Source
View on FDA website ↗

Reason for Recall

ALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.

Distribution

Worldwide

Lot / Code Info

Serial Numbers 2975 to 5360

Root Cause

Other

Action Taken

On May 24, 2006 the manufacturer initiated this recall. The Recalling establishment/US Initial Importer/Distributor-received the manufacturer''s notice on May 31, 2006 and initiated their recall on June 1, 2006. The US distributor has issued a Medical Device Field Correction.