Access Immunoassay system CEA, in vitro diagnostic, Part No. 33200
This recall has been terminated (originally issued August 1, 2006).
- Company
- Beckman Coulter Inc.
- Recall Initiated
- April 12, 2006
- Posted
- August 1, 2006
- Terminated
- April 3, 2012
- Recall Number
- Z-1313-06
- Quantity
- 775
- Firm Location
- Brea, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Firm has confirmed that the Access CEA Assay may generate decreased values with certain frozen platelet specimens.
Distribution
Nationwide and Canada
Lot / Code Info
All codes
Root Cause
Other
Action Taken
A Product Corrective Action letter was sent via US mail the week of April 10, 2006 to all Access CEA customers informing them that the Access CEA Assay could generate decreased values with certain frozen patient specimens. Although they have found it at a low occurrence, it is recommended that specimens should not be frozen prior to analysis until further notice. The letter informs customers that Beckman Coulter has verified sample stability through at least 14 days when properly stored at 2-8 degrees Celsius and shipping stability for 4 days at 2-8 degrees Celsius using cool packs. Firm also advises customers to contact customer technical support or local Beckman Coulter Representative if laboratory is unable to avoid freezing specimens and to review results from frozen CEA samples to ensure that the match the total clinical presentation per the customer''s established laboratory procedure. The letter requests that the notification remain part of the laboratory Quality System documentation and that the response form be returned to the firm within 10 days.