The device's marketing brochure states: "IntelliVue M3150 Information Center...Philips Medical Systems". The beige-colored PAS-210 external speakers can be identified by their use of DB15-to-3.5 mm mini-phone jack cable (Agispec-081Y-4 or Agispec-081-6) to connect to the rp5700 PC. Indicated for central monitoring of multiple adult, pediatric, and neonatal patients.
This recall has been terminated (originally issued October 8, 2010).
- Recall Initiated
- November 21, 2008
- Posted
- October 8, 2010
- Terminated
- October 12, 2010
- Recall Number
- Z-0034-2011
- Quantity
- 424 units in US.
- Firm Location
- Bothell, WA
- Official Source
- View on FDA website ↗
Reason for Recall
Potential for Internal Sound Amplifier to malfunction which could delay treatment.
Distribution
Nationwide Distribution
Lot / Code Info
Serial numbers are: 4810A03287, 4810A03260, 4810A03261, 4810A03262, 4810A04902, 4810A04906, 4810A04907, 4810A04908, 4810A04909, 4810A04910, 4810A04911, 4810A04912, 4810A05422, 4810A05423, 4810A11902, 4810A03367, 4810A03368, 4810A03369, 4810A03370, 4810A03372, 4810A03231, 4810A03240, 4810A03241, 4810A03242, 4810A03254, 4810A03255, 4810A03256, 4810A03257, 4810A03243, 4810A03244, 4810A03245, 4810A02772, 4810A04096, 4810A04097, 4810A04094, 4810A04095, 4810A04084, 4810A04085, 4810A03710, 4810A03711, 4810A03407, 4810A03408, 4810A03409, 4810A03410, 4810A04494, 4810A03934, 4810A03935, 4810A04768, 4810A04783, 4810A04809, 4810A04810, 4810A03631, 4810A04805, 4810A04598, 4810A11904, 4810A04053, 4810A11873, 4810A05158, 4810A05167, 4810A03078, 4810A03762, 4810A11903, 4810A05378, 4810A04798, 4810A04799, 4810A04800, 4810A04764, 4810A04765, 4810A04801, 4810A04802, 4810A04803, 4810A04804, 4810A04784, 4810A04785, 4810A04786, 4820A10136, 4820A10138, 4820A10139, 4820A10124, 4820A10125, 4820A11906, 4820A11905, 4810A03623, 4810A04796, 4810A04763, 4810A04777, 4810A04778, 4810A04779, 4810A04780, 4810A04781, 4810A09398, 4810A11851, 4810A03643, 4810A04664, 4810A04650, 4810A04651, 4810A04652, 4810A09748, 4810A09749, 4810A04565, 4810A04567, 4810A04568, 4810A05277, 4810A05278, 4810A05280, 4810A05281, 4810A04437, 4810A04438, 4810A05288, 4810A05289, 4810A05290, 4810A05291, 4810A05292, 4810A05293, 4810A05282, 4810A05283, 4810A05284, 4810A05285, 4810A05286, 4810A05287, 4810A04573, 4810A04574, 4810A04663, 4810A02713, 4810A02714, 4810A02715, 4810A02716, 4810A02762, 4810A02770, 4810A05397, 4810A05398, 4810A05399, 4810A05400, 4810A11881, 4810A05424, 4810A09648, 4810A09649, 4810A09650, 4810A09651, 4810A03699, 4810A03700, 4810A03702, 4810A03703, 4810A02941, 4810A02942, 4810A02943, 4810A02944, 4810A02945, 4810A02946, 4810A02947, 4810A02948, 4810A09687, 4810A11858, 4810A11877, 4810A02976, 4810A02962, 4810A02963, 4810A02964, 4810A02965, 4810A02966, 4810A02922, 4810A05425, 4810A05413, 4810A05417, 4810A04557, 4810A04555, 4810A04556, 4810A04504, 4810A04505, 4810A04506, 4810A04507, 4810A04508, 4810A04509, 4810A04510, 4810A04511, 4825A06062, 4825A06065, 4825A06063, 4825A06064, 4810A09047, 4810A09048, 4810A09049, 4810A09050, 4810A09051, 4810A09052, 4810A09053, 4810A09054, 4810A09055, 4810A09056, 4810A09084, 4810A09085, 4810A09086, 4810A09087, 4810A09302, 4815A05570, 4815A05571, 4815A05574, 4815A05573, 4810A05247, 4810A09756, 4810A11893, 4810A11894, 4810A10015, 4810A10016, 4810A10018, 4810A10019, 4810A10020, 4810A10021, 4810A02915, 4810A02874, 4825A06101, 4825A06102, 4825A06116, 4825A06117, 4810A09436, 4810A10078, 4810A10079, 4810A10083, 4810A09634, 4810A09635, 4810A09654, 4810A09655, 4810A09656, 4810A09657, 4810A09640, 4810A09645, 4810A09658, 4810A09641, 4810A09642, 4810A09659, 4810A09660, 4810A09661, 4810A09662, 4810A09463, 4810A09425, 4810A09462, 4810A09464, 4810A09892, 4810A09894, 4810A09876, 4810A09877, 4810A09878, 4810A09879, 4810A09880, 4810A09881, 4810A09895, 4810A09893, 4810A09882, 4810A09883, 4810A09884, 4810A05147, 4810A04228, 4810A04292, 4810A04293, 4810A04294, 4815A05914, 4815A05915, 4815A05882, 4815A05883, 4815A05884, 4815A05901, 4815A05902, 4815A05903, 4815A05904, 4815A05905, 4815A05906, 4810A02786, 4810A02787, 4810A02788, 4825A06053, 4825A06049, 4825A06050, 4825A06051, 4825A06052, 4825A06044, 4825A06045, 4825A06046, 4825A06047, 4825A06048, 4825A06092, 4825A06038, 4825A06039, 4825A06040, 4825A06041, 4825A06042, 4825A06043, 4825A06056, 4810A04970, 4810A04971, 4810A04974, 4810A04975, 4810A04976, 4810A04977, 4810A04978, 4810A04979, 4810A05030, 4810A05031, 4810A03323, 4810A03324, 4810A03325, 4810A09359, 4810A09360, 4810A05083, 4810A04544, 4810A04545, 4810A04546, 4810A04547, 4810A04548, 4810A04550, 4810A04551, 4810A04552, 4810A04553, 4810A09215, 4810A09199, 4810A09200, 4810A09201, 4810A09202, 4810A09203, 4810A09212, 4810A04558, 4810A04559, 4810A04560, 4810A04561, 4810A04562, 4810A04563, 4810A04564, 4810A04048, 4810A04049, 4810A04050, 4810A04051, 4810A11857, 4810A10087, 4810A10088, 4810A10089, 4810A10090, 4810A10091, 4810A10092, 4810A10093, 4810A10094, 4815A05948, 4815A05949, 4815A05950, 4815A05951, 4815A05895, 4815A05896, 4815A05897, 4815A05898, 4810A04812, 4810A04813, 4810A03652, 4810A03653, 4810A03839, 4810A11896, 4810A11897, 4810A03575, 4810A03576, 4810A03577, 4810A03578, 4810A03581, 4810A03582, 4810A03583, 4810A03584, 4810A05406, 4810A05407, 4810A09574, 4810A09575, 4810A09576, 4810A09577, 4810A09578, 4810A09579, 4810A04662, 4810A04673, 4810A04674, 4810A03206, 4810A04303, 4810A04304, 4810A04305, 4810A04306, 4810A04386, 4810A05459, 4810A05460, 4810A05461, 4810A05504, 4815A05813, 4815A05814, 4815A05818, 4810A04124, 4810A04125, 4810A04126, 4810A04127, 4810A04128, 4810A04129, 4810A04352, 4810A04353, 4810A04354, 4810A04355, 4815A05604, 4815A05617, 4815A05618, 4815A05619, 4815A05634, 4810A04980, 4810A04998, 4810A04999, 4810A05000, 4810A05001, 4810A05002, 4810A03229, 4810A03205, 4810A03249, 4810A03204, 4810A03230, 4810A03250, 4810A05024, 4810A04981, 4810A04982, 4810A04983, 4810A04984, 4810A05025, 4810A05026, 4810A05027, 4810A05028, 4810A05009, and 4810A05012.
Root Cause
Component change control
Action Taken
On 11/20/08, Philips began sending the URGENT-Medical Device Correction letter to their consignees via e-mails. On 11/21/08, the firm sent out the recall letter via United State Postal Service. The letter states the affected device as Philips IntelliVue Information Centers, running on the HP rp5700 PC platform equipped with an active speaker solutions using a DB-15 to 3.5 mm cable set. The consignees are advised to ensure that the audible alarm system is clear, free of noise and distortion and adjustable to different sound levels which is explained in p. 5-47 "Adjusting Alarm Tone Volume" of IntelliVue of the Instruction for Use, Release K Manual, supplied with the hardware, Philips Part Number 453564062301. Consignees can call Philips Healthcare Customer Care Service Center at 1-800-722-9377, #3, #1 and reference FCO 99862012.