GE Healthcare Innova 4100 Digital Fluoroscopic Imaging (Cardiovascular Imaging System

This recall has been terminated (originally issued September 30, 2008).

Company
Datex Ohmeda, Inc
Recall Initiated
May 19, 2008
Posted
September 30, 2008
Terminated
June 30, 2016
Recall Number
Z-1522-2008
Quantity
87
Firm Location
Madison, WI
Official Source
View on FDA website ↗

Reason for Recall

GE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located in the control room).

Distribution

Worldwide Distribution - AL, AZ, CA, FL,, GA, HI, IA,IL, IN, KY, LA,MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, and WV, Puerto Rico. OUS to include: Australia, Byelarus, China, Denmark, France, Germany, Greece, Honduras, Iran, Iraq, Italy, Japan, Latvia, Lebanon, Mexico, Poland, Russia, Saudi Arabia, Turkey, United Kingdom, and Venezuela.

Lot / Code Info

Systems id: 77573841, 80158241, 910063039, 3018774100, 4103374100, 4438494100, 5134754100, 5138534100, 6029234100, 8159374100, 8595724100, 9408984100, 82406010011, 082416130011, 82416130012, 00095VAS07, 205759CL2, 216844INNOVA, 229228INNOV41V, 229434INNOV, 239436CATH1, 251435ICV5, 304255INNOVA, 304347INNOVA, 315448RM4, 315448RM5, 319338VA4100, 325670CATH, 337981INNOVA, 352333VASC, 352376SP, 352401CATH, 352401VASC, 407518SP2, 407518SP3, 407SPECIALS, 414961INN4100, 478633INOV, 508856OR41, 515263IL4100, 516562INNOVA4, 573632CRA, 573761SMA, 574535IN41, 600050VA01, 6012884100SP1, 606833INNOVA, 609652CATH3, 614293LIBERTY1, 614293LIBERTY2, 704487CR41, 706802FINN, 715369INN4100, 717763INNOVA41, 717763OR4100, 718518CVL1, 724656ANG, 727725SPEC1, 727819SPSC, 740592INNOVA, 757398INNOVA, 757736INNOVA, 803329CL41, 804320CATH2B, 804594CATH, 812279INNOVA1, 813972INTV, 816276SP1, 817255NH4100, 828696IN41, 865541CCL5, 865693CCL3, 901726SP2, 903CSM4100, 904276SP, 906485ANGIO, 918307SHIN4, 941483INNOVA41, 956632INN4100, A5630803, B5801711, BE5005VA01, BY4020VA01, RU1018VA03, SA1188VA03, SLIN4100, YV0003.

Root Cause

Software design

Action Taken

An Urgent Medical Device Correction letter, dated June 2008 was sent to customers The letter described the issue, affected products and provided short term user recommendations. GE developed a new software release that corrects this condition and will launch the corrective action to upgrade all affected systems.