Acu-Loc Distal Radius Plate, Standard, Left. Product labeled in part, " Acu-Loc Distal Radius Plate, Standard, Left... PL-DR50L W48003... Manufactured or Distributed by: Acumed LLC Hillsboro, OR ".

This recall has been terminated (originally issued August 28, 2008).

Company
Acumed Llc
Recall Initiated
June 29, 2007
Posted
August 28, 2008
Terminated
February 3, 2010
Recall Number
Z-2153-2008
Quantity
51 pieces
Firm Location
Hillsboro, OR
Official Source
View on FDA website ↗

Reason for Recall

Plates are incorrectly laser marked. The PL-DR50L plates are incorrectly laser marked with PL-DR70L.

Distribution

Worldwide Distribution --- including USA and countries of South Africa, Italy, Finland, and Brazil.

Lot / Code Info

Lot: W4800; Expiration date: 2012-02

Root Cause

Process control

Action Taken

The firm issued an Urgent Notification: Product Removal letter on 6-29-2007. The notification was reportedly sent by either FAX or e-mail. The notification informed consignees to remove the plates from their inventory and return them to Acumed. The notification did not include any instructions for a sub-recall. On 6/13/08, the firm sent a second Urgent Notification: Product Removal letter. This letter describes the product as PL-DR50L - Lot Code - W48003, explains the problem and asks customers to return product to Acumed.