RecallDepth

Left Locking Proximal Metatarsal Wedge Plate. Product labeled in part, "ACUMED 5885 NW Cornelius pass Road Hillsboro, OR. 97124-9432... Size: Left Locking Proximal Metatarsal Wedge Plate... REF 70-0010-S... LOT W58118".

This recall has been terminated (originally issued September 20, 2008).

Company
Acumed Llc
Recall Initiated
February 14, 2008
Posted
September 20, 2008
Terminated
February 3, 2010
Recall Number
Z-2111-2008
Quantity
3 pcs
Firm Location
Hillsboro, OR
Official Source
View on FDA website ↗

Reason for Recall

Product sterility may be compromised.

Distribution

Worldwide Distribution ---- USA including states of WA, NE, NC, UT, FL, CA, OR, TX, NY, IL, OH, MT, and Puerto Rico, and countries of United Kingdom, Australia, and Greece.

Lot / Code Info

Work order number/Lot: W58118 (W=work order, ###=sequential number); Expiration date: 2014-11 (Year/Month)

Root Cause

Packaging process control

Action Taken

The firm issued four separate Urgent Notification: Product Removal were issued by the firm on 2-14-2008, 2-20-2008, 2-26-2008, and 3-25-2008 for the issue of poorly sealed sterile pouches. The notifications were reportedly sent by either FAX or e-mail. The firm explained the issue and requested customers to return the specified products. On 7/16/08, the firm issued a second recall notification to their customers. This letter informed the customer of the recall, listed the products affected, stated the reason for the recall, tells customers, "return existing inventories of the recalled products to Acumed... If you further distributed this product, please notify those customers/consignees of this recall. This recall is to be extended to the user level." The letter includes an effectiveness form to be filled out by the customer and returned to Acumed.