RecallDepth

CIBASoft Visitint Soft Contact Lenses, SKU #749591001385, Power -6.50, Diameter 13.8, Base Curve, 8.9, Expiration 2011-10.

Company
Ciba Vision Corporation
Recall Initiated
April 30, 2007
Posted
May 9, 2007
Terminated
March 13, 2008
Recall Number
Z-0793-2007
Quantity
22 lenses
Firm Location
Duluth, GA

Reason for Recall

Base curve of the lenses do not meet specification.

Distribution

Nationwide.

Lot / Code Info

Lot #41074100201110

Root Cause

Other

Action Taken

Consignees were contacted via traceable mail on 4/30/2007. The initial notification was sent to Eye Care Professionals. The firm subsequently identified one distributor and one retailer. The recall letter was revised and sent to these consignees on May 2, 2007. Specific return instructions were attached along with a Recall Tracking Form that was to be completed and returned to CIBA Vision. They were instructed to return any of the affected product on hand to CIBA Vision for replacement.

More recalls by Ciba Vision Corporation