Gyrus ACMI Tripolar Ultra Cutting Forceps: 10 mm X 32 cm-Reference Number: 006689-901

This recall has been terminated (originally issued March 13, 2007).

Company
Gyrus Acmi Corporation
Recall Initiated
February 19, 2007
Posted
March 13, 2007
Terminated
February 24, 2008
Recall Number
Z-0614-2007
Quantity
184 boxes (5 units/box)
Firm Location
Southborough, MA
Official Source
View on FDA website ↗

Reason for Recall

Product sterility may be compromised due to failure of the tray seal packaging

Distribution

Worldwide distribution ----USA and countries of Canada, The Netherlands, Japan, Portugal, Columbia, South Africa, Australia, and Spain.

Lot / Code Info

Lot Nos.: 6258014, 6303028, 6306017, 6312602, 6314021, 6320034, 6326096, 6335040, 6342010, 6362025, 7004015, 7009015

Root Cause

Packaging process control

Action Taken

Gyrus ACMi issued recall letters on 2/19/07. Each account is asked to remove recalled product and return product .