HTA Console System, 230 Volt , French, Product Number: 56001FO (HTA system is a software-controlled hysteroscopic thermal endometrial ablation system)

Company
Boston Scientific Corporation One Boston
Recall Initiated
January 24, 2007
Posted
March 31, 2007
Terminated
December 12, 2007
Recall Number
Z-0690-2007
Quantity
2 units
Firm Location
Scientific Place Natick, MA

Reason for Recall

Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment

Distribution

Worldwide-USA, UK, Denmark, Italy, and France

Lot / Code Info

Unit Numbers: IN0350JG, IN0351JG,

Root Cause

Other

Action Taken

Boston Scientific notiifed accounts by letter dated January 24, 2007 via federal Express Mail. Accounts were requested to examine inventory, remove and return product.