Boston Scientific Cutting Balloon Ultra2 Monorail Device, Boston Scientific, One Boston Scientific Place, Natick, MA 01760.

This recall has been terminated (originally issued September 23, 2006).

Company
Boston Scientific
Recall Initiated
July 21, 2006
Posted
September 23, 2006
Terminated
February 3, 2007
Recall Number
Z-1512-06
Quantity
1,000
Firm Location
Maple Grove, MN
Official Source
View on FDA website ↗

Reason for Recall

Lack of assurance of sterility (pre-sterilization bioburden limits exceeded)

Distribution

Class II Recall - Nationwide distribution ---- including states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV, and WY.

Lot / Code Info

Material Number (UPN) / Catalog Number / Lot/Batch #:  H749CBM200060 / CBM20006 / EG0438 & EG0499;  H749CBM200100 / CBM20010 / EG0427;  H749CBM225060 / CBM22506 / EG0448 & EG0471;  H749CBM225100 / CBM22510 / EG0444 & EG0502;  H749CBM250060 / CBM25006 / EG0439 & EG0472;  H749CBM250100 / CBM25010 / EGO425, EG0440, EG0474 & EG0475;  H749CBM250150 / CBM25015 / EG0466;  H749CBM275060 / CBM27506 / EG0420 & EG0487;  H749CBM275100 / CBM27510 / EG0419 & EG0456;  H749CBM275150 / CBM27515 / EG0424;  H749CBM300060 / CBM30006 / EG0423, EG0473 & EG0497;  H749CBM300100 / CBM30010 / EG0426, EG0441, EG0465, EG0485;  H749CBM300150 / CBM30015 /EG0443 & EG0418;  H749CBM325060 / CBM32506 / EG0500;  H749CBM325100 / CBM32510 / EG0422 & EG0501;  H749CBM325150 / CBM32515 / EG0467;  H749CBM350060 / CBM35006 / EG0437;  H749CBM350100 / CBM35010 / EG0349, EG0428, EG0468;  H749CBM350150 / CBM35015 / EG0421 & EG0486;  H749CBM375060 / CBM37506 / EG0503;  H749CBM375100 / CBM37510 / EG0457;  H749CBM375150 / CBM37515 / EG0455;  H749CBM400060 / CBM40006 / EG0504;  H749CBM400100 / CBM40010 / EG0442 & EG0498;  H749CBM400150 / CBM40015 / EG0476.

Root Cause

Other

Action Taken

On July 21, 2006, Boston Scientific initiated a voluntary recall of 45 lots/batches of Cutting Balloon Ultra Monorail Devices due to their determination that the lots do not meet their sterilization requirements. There is a remote possibility of infection in the patient. An infection, if it occurred, could be recognized and treated effectively in these patients. Boston Scientific is not aware of any complaints related to this product issue. Letters dated July 21, 2006, were sent to their customers with instructions to cease further distribution, complete and return a tracking form, and product return instructions.