Stat Profile Critical Care Xpress CCX+ Analyzer Catalog Number: 37413
- Company
- Nova Biomedical Corporation
- Recall Initiated
- March 30, 2004
- Posted
- July 20, 2004
- Terminated
- June 17, 2004
- Recall Number
- Z-0827-04
- Quantity
- 30 units
- Firm Location
- Waltham, MA
Reason for Recall
Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems
Distribution
CA, IL, LA, MA ,NJ, NY, OH, PA, TX International: Italy ,England, Japan,Turkey., U.A.E., Greece, Thailand, Switzerland, Finland, Saudi Arabia, Italy, Mexico, Trinidad, Ireland, France, Cyprus, Czech Republic, ,Germany
Lot / Code Info
Domestic Serial Numbers: Y02C03170 Y02C03140 Y02C03180 Y02104060 Y02104180 Y02104070 Y02104210 Y02104080 Y02204180 International S/N: Y02204030 Y02204040 Y02204050 Y02204110 Y02204120 Y02204140 Y02204190 Y02304020 Y02C03190 Y02C03160 Y02C03090 Y02204060 Y02C03070 Y02304050 Y02C03050 Y02104090 Y02104230 Y02104170 Y02104020 Y02304070
Root Cause
Other
Action Taken
Nova Biomedical notified users by telephone on 3/31/04 and conducted follow-up by letter either by email or fax. On-site visit will be scheduled to replace the pump assembly.