NucliSensEasy Q Incubator, Product Number 285204 (this number can be found at the back of the instrument).
This recall has been terminated (originally issued July 20, 2004).
- Company
- Biomerieux, Inc.
- Recall Initiated
- February 26, 2004
- Posted
- July 20, 2004
- Terminated
- March 24, 2004
- Recall Number
- Z-0735-04
- Quantity
- 3 units
- Firm Location
- Durham, NC
- Official Source
- View on FDA website ↗
Reason for Recall
Instrument malfunction. Electrical short circuit may occur resulting in damage or a fire hazard.
Distribution
CT, MD, MT
Lot / Code Info
Catalog (or Product) number 285204, Serial Number 36061, 36065 and 36068
Root Cause
Other
Action Taken
This correction is being handled as an instrument repair (rework) by bioMerieux Field Service; therefore no communication has been sent to the customer sites. The repair involves heat sink replacement or use of insulating tape on the rear side of the keypad.