The device is SCIPRO Flexible Stone Dislodger and Retrieval Set, Catalog # SP3000146-ST used for removal of calculi, fragments of medical devices and other foreign objects uring urological or gatroenterological procedures. The device has the manufacturer listed on the label which is Scion Cardio-Vascular, Inc., 14256 SW 119 Avenue, Miami, FL 33186.

Company
Scion Cardio-Vascular, Inc.
Recall Initiated
June 30, 2003
Posted
February 27, 2004
Terminated
May 7, 2004
Recall Number
Z-0557-04
Quantity
41
Firm Location
Miami, FL

Reason for Recall

The firm received field complaints that it was difficult to open the stone retrieval basket during use.

Distribution

The software systems were installed at 2 medical centers in Illinois and Alabama.

Lot / Code Info

Lot Number OML30522-01

Root Cause

Other

Action Taken

The firm mailed a recall letter to there consignees on June 30, 2003 requesting return of any unused devices. The firm has received 100% response and accounted for all recalled devices.