StatLock CV Plus w/Pigtail, Catheter Stabilization Device Sterile EO Single Use Only, C.R. Bard, Inc. The StatLock device provides stabilization for compatible medical tubes and catheters.

This recall has been terminated (originally issued June 28, 2013).

Company
Bard Access Systems
Recall Initiated
June 10, 2013
Posted
June 28, 2013
Terminated
May 21, 2014
Recall Number
Z-1618-2013
Quantity
3,875 units.
Firm Location
West Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

BARD Access Systems is recalling lot JUWJF283 of StatLock CV Plus Stabilization Butterfly Devices due to incorrect pad assembly used during manufacture. The absence of the hydrocolloid strip on the bottom may pose an incremental risk including localized skin irritation. Missing directional arrow may lead to mis-orientation of the StatLock stabilization which may impact the securement of the cathet

Distribution

National and Global distribution (Europe, European Economic Area, Argentina, China, and Brazil).

Lot / Code Info

Lot JUWJF283

Root Cause

Packaging

Action Taken

On 6/18/13, an "Urgent Kit Recall" notification letter was mailed to all customers informing them of the recall and instructing the consignee to remove and return any affected product lot numbers to Bard Access Systems through return goods authorization. Customers may contact 800-290-1689 if questions.