Cordis OPTEASE Retrieval Inferior Vena Cava (IVC) Filter***REF Cat. No.***466F220A and 466F220B***Vena Cava Filter and Introduction Kit***(BRITE TIP Catheter Sheath Introducer and Angiographic Vessel Dilator)***Cordis***a Johnson and Johnson Company***ENDOVASCULAR*** Label No. SB 10247519.2/***LB466F220AO.5***omoraoro/2012.050794***Do not resterilize***. Product Usage: The OPTEASE Vena Cava Filter is indicated for use in the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the vena cava in several situations: Pulmonary thromboembolism when anticoagulants are contraindicated; Failure of anticoagulant therapy for thromboembolic disease; Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced and Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.

This recall has been terminated (originally issued August 7, 2013).

Company
Cordis Corporation
Recall Initiated
March 29, 2013
Posted
August 7, 2013
Terminated
October 27, 2014
Recall Number
Z-1887-2013
Quantity
33,000 units
Firm Location
Miami Lakes, FL
Official Source
View on FDA website ↗

Reason for Recall

Correction to provide clarification and additional labeling to minimize likelihood on implanting filter backwards.

Distribution

Worldwide Distribution - USA Nationwide in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, Ny, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, including Hawaii and Puerto Rico and the countries of Canada & Panama

Lot / Code Info

Model Numbers 466F220A and 466F220B.

Root Cause

Labeling design

Action Taken

Cordis Corporation sent an "URGENT Medical Device Correction" notification to their consignees in the US by letter dated April 3, 2013 and an e-mail was sent to the Canadian consignees. The letter identified the affected product, problem and actions to be taken. Consignees were instructed to read the description and recommendations sections carefully, then sign and return the enclosed Acknowledgement Form directly to Cordis to their fax number(908) 429-8287 or Scan and e-mail to MiamiQA@crdus.jnj.com. The letter states to pass on the notice to anyone in the consignees' facilities that need to be informed. Additionally, the notification request consignees to maintain awareness of this notification until the information has been incorporated in the Cordis OPTEASE Vena Cava Filter labeling. For questions related to the issue contact Cordis Endovascular Medical Information Line 877-338-4235.