Smiths Medical CADD-Sentry Pro Medication Safety Software - electronic pump protocol programming software - Administrator REF 21-6275-51 RX only, Smiths Medical MD, Inc. St Paul, Minnesota 55112 USA

Company
Smiths Medical Md, Inc.
Recall Initiated
August 14, 2007
Posted
February 26, 2008
Terminated
October 31, 2008
Recall Number
Z-0134-2008
Quantity
3 CD-ROMs
Firm Location
Saint Paul, MN

Reason for Recall

Incorrect dosage possible: A software anomaly in the CADD-Sentry Pro Administrator software (version B) where the values of a weight based protocol go to zero or round up to the next higher hundredths place, even if Administrator displayed the values correctly at data entry.

Distribution

UT, MD, CO

Lot / Code Info

Catalog Number: 21-6275-51, Software Version B.

Root Cause

Software design

Action Taken

A Smiths Medical MD employee visited the consignees between August 14 and 31, 2007, to verbally notify them of the recall, collect the recall CD-ROM containing the affected software and replace with the updated software.