Smiths Medical CADD-Sentry Pro Medication Safety Software - electronic pump protocol programming software - Administrator REF 21-6275-51 RX only, Smiths Medical MD, Inc. St Paul, Minnesota 55112 USA

This recall has been terminated (originally issued February 26, 2008).

Company
Smiths Medical Md, Inc.
Recall Initiated
August 14, 2007
Posted
February 26, 2008
Terminated
October 31, 2008
Recall Number
Z-0134-2008
Quantity
3 CD-ROMs
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

Incorrect dosage possible: A software anomaly in the CADD-Sentry Pro Administrator software (version B) where the values of a weight based protocol go to zero or round up to the next higher hundredths place, even if Administrator displayed the values correctly at data entry.

Distribution

UT, MD, CO

Lot / Code Info

Catalog Number: 21-6275-51, Software Version B.

Root Cause

Software design

Action Taken

A Smiths Medical MD employee visited the consignees between August 14 and 31, 2007, to verbally notify them of the recall, collect the recall CD-ROM containing the affected software and replace with the updated software.