SYNCHRON Systems Vancomycin Reagent, in vitro diagnostic. Part Number 474824.

This recall has been terminated (originally issued May 18, 2006).

Company
Beckman Coulter Inc.
Recall Initiated
March 30, 2006
Posted
May 18, 2006
Terminated
February 29, 2012
Recall Number
Z-0886-06
Quantity
3147
Firm Location
Brea, CA
Official Source
View on FDA website ↗

Reason for Recall

Beckman Coulter has received reports of patient samples and quality control recovery shifting low when using Vancomycin reagent lots M507219 and M511327. Internal QC recovery was as much as 15% lower than the assigned target value.

Distribution

Nationwide & Canada

Lot / Code Info

Lot numbers M507219 & M511327

Root Cause

Other

Action Taken

A Product Corrective Action (PCA) letter was be sent the week of MAR 30, 2006 to all SYNCHRON Systems Vancomycin Reagent customers of lot numbers M507219 and M511327 informing them to discontinue use and discard reagent. (Letters were sent by US mail)