RecallDepth

ETI-MP IgM Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of specific IgM antibodies to Mycoplasma pneumoniae in human serum. Test kit for 192 determinations. Catalog No. 4700. For In Vitro Diagnostic use. manufactured for DiaSorin Inc., 1950 Northwestern Avenue P.O. Box 285, Stillwater, MN 55082-0285

Company
Diasorin Inc.
Recall Initiated
October 20, 2004
Posted
March 18, 2005
Terminated
July 22, 2006
Recall Number
Z-0633-05
Firm Location
Stillwater, MN

Reason for Recall

2 lots of ETI-MP IgM assay kits showed a decline in Optical Density (OD) reading of the positive control and calibrators, P10 and P50. The decline in OD of the positive control and calibrators (P10 & P50) would cause the assay validation criteria to fail.

Distribution

MA, OH & IL

Lot / Code Info

lot # E052660D, E052790D

Root Cause

Other

Action Taken

Customers were sent notification letters. Recommended to detroy the remaining kits and requested a form be returned which indicates the number of kits destroyed so they can arrange replacement.

More recalls by Diasorin Inc.