RecallDepth

Solar 9500 Information Monitor with software version 3A.

This recall has been terminated (originally issued December 18, 2003).

Company
General Electric Medical Systems Information Technology
Recall Initiated
November 10, 2003
Posted
December 18, 2003
Terminated
November 23, 2005
Recall Number
Z-0250-04
Quantity
maximum 648 Solar 9500s with Version 3A software (297 Solar 9500 Vers. 3A software were manufactured plus 131 software update kits to update multiple)
Firm Location
Milwaukee, WI
Official Source
View on FDA website ↗

Reason for Recall

Software gives inaccurate readings when transferring TRAM module from Solar 9500 to other Solar monitoring systems without discharging TRAM first.

Distribution

The product was distributed to consignees located nationwide in the United States and worldwide.

Lot / Code Info

Solar 9500 Information Monitor with software version 3A.

Root Cause

Other

Action Taken

A letter dated November 10, 2003 was sent to the consignees. The letter recommended certain actions to avoid the problems and provided options for the software to be updated.