RecallDepth

TI Standard Polyaxial Screw Assembly 6.5MM x 55MM. Part # 62065-55. Subsequent Product Codes: MNH, MNI. The Zodiac Polyaxial Pedicle Screw System is intended for use as a posterior spinal fixation device to aid in the surgical correction of various spinal deformities and pathologies in the thoraco-lumbo-sacral portion of the spine. It intended to provide stabilization during the development of fusion utilizing a bone graft. Specific indications for the Zodiac Polyaxial Pedicle Screw System are dependent in part on the configuration of the assembled device and the method of attachment to the spine.

This recall has been terminated (originally issued November 8, 2012).

Company
Alphatec Spine, Inc.
Recall Initiated
December 9, 2010
Posted
November 8, 2012
Terminated
November 9, 2012
Recall Number
Z-0269-2013
Quantity
11 units
Firm Location
Carlsbad, CA
Official Source
View on FDA website ↗

Reason for Recall

The recall was initiated because Alphatec Spine confirmed that the TI Standard Polyaxial Screw Assembly was laser marked with the incorrect length of 45 MM.

Distribution

Nationwide Distribution including TN, WI, and FL.

Lot / Code Info

Lot # 634745

Root Cause

Error in labeling

Action Taken

Alphatec Spine distributors who received the TI Standard Polyaxial Screw Assembly were contacted by phone and by email on December 9, 2010. A phone script was used as a recall notification to contact their three distributors who received the recalled products. The recall notification informed the customers of the problem identified and the action to be taken. Customers were informed that they will receive replacements within 2-3 days. Customers with questions were instructed to call (800) 922-1356. For questions regarding this recall call 760-431-9286.

More recalls by Alphatec Spine, Inc.

View all recalls by this company →