RecallDepth

Perkin Elmer Genetic Screening Processor (GSP) Product code: 2021-0010. It is intended for in vitro quantitative / qualitative determination of analytes in body fluids.

This recall has been terminated (originally issued February 28, 2012).

Company
Perkinelmer Life And Analytical Sciences, Wallac, Oy Mustionkatu 6
Recall Initiated
January 9, 2012
Posted
February 28, 2012
Terminated
March 6, 2013
Recall Number
Z-1116-2012
Quantity
10 units
Firm Location
Turku Finland
Official Source
View on FDA website ↗

Reason for Recall

GSP instruments with defective washer modules may cause a false negative screening result being reported for T4, TSH and IRT. For 17-OHP, a false positive result may be generated if the failure occurs for a normal sample.

Distribution

Worldwide Distribution - USA, including the states of Florida, Ohio, Oregon, Texas and the countries of Brazil, Belgium, France, Germany, and the UK.

Lot / Code Info

Serial numbers from 20210012 to 20210021

Root Cause

Nonconforming Material/Component

Action Taken

Perkin Elmer issued an "Urgent Medical Device Recall" letter dated 1/25/12 to all customers and informed them about the possible issue and requested permission for service personnel to check their instrument. The customers were instructed that the instrument could not be used and the washer module required replacement. Customers were told to expect to be contacted by the firm's local representative for further details.

More recalls by Perkinelmer Life And Analytical Sciences, Wallac, Oy Mustionkatu 6

View all recalls by this company →