DePuy Spine VIPER POLY SCREW 5.5 8 X80MM Tl Product Code: 1867-15-880 Spinal Implant Component.

Company
Depuy Spine, Inc.
Recall Initiated
November 7, 2012
Posted
January 24, 2013
Terminated
December 24, 2013
Recall Number
Z-0706-2013
Quantity
36 units
Firm Location
Raynham, MA

Reason for Recall

Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head

Distribution

Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.

Lot / Code Info

All lots

Root Cause

Component change control

Action Taken

DePuy Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated November 7, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory, quarantine the material and return remaining inventory. Customers were advised to complete the enclosed Business Reply Form and fax back to 508-828-3762 or email to lregan3@its.jnj.com. For questions call Linda Regan @508-977-6606.