BD GeneOhm Cdiff Assay , Catalog # 441401, 200 box tests 2-3 labeled in part***GeneOhm Sciences Canada, Inc., 2555 boul, du Parc-Technologique, Quebec, QC, Canada G1P 4S5*** BD GeneOhm Cdiff Assay is a qualitative in vitro diagnostic test for the direct detection of C. difficile toxin gene (tcdB) in human liquid or soft stool specimens from patients suspected of having Clostridium difficile-associated disease (CDAD). The test, based on real-time PCR, is intended for use as an aid in diagnosis of CDAD. The test is performed directly on the specimen, utilizing polymerase chain reaction (PCR) for the amplification of specific targets and fluorogenic target-specific hybridization probes for the detection of the amplified DNA.
This recall has been terminated (originally issued September 6, 2012).
- Recall Initiated
- May 22, 2012
- Posted
- September 6, 2012
- Terminated
- November 2, 2012
- Recall Number
- Z-2319-2012
- Quantity
- 32 kits
- Firm Location
- Sparks, MD
- Official Source
- View on FDA website ↗
Reason for Recall
Leakage in Cepheid SmartCycler reaction tubes.
Distribution
Nationwide Distribution including CA, DE, FL, GA, IL, IN, KS, KY, ME, MT, NC, NJ, NM, NY, OH, OK, PA, SC, TX, VA, WA, WI, and WV.
Lot / Code Info
Lot/ Exp. 08T11250 2012-07-06
Root Cause
Nonconforming Material/Component
Action Taken
BD Diagnostic Systems issued an Urgent Product Recall letter via UPS overnight delivery on May 22, 2012. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to take the following actions: Discontinue use and discard any affected product. BD would send replacement tubes for the discarded material. Complete the attached form whether or not they have any inventory remaining so that the firm may acknowledge their receipt of the notification. Fax the form to: Attention: Regulatory Compliance FAX: 410-316-4258 Email: RegulatoryComplianceFax@bd.com If further assistance is needed, customers should contact BD Technical Services Department at 1-800-838-8663. For questions regarding this recall call 410-316-4054.