RecallDepth

Allura X-ray System. The Allura X-ray System is an angiographic X-ray system.

This recall has been terminated (originally issued April 25, 2012).

Company
Philips Healthcare Inc.
Recall Initiated
April 2, 2012
Posted
April 25, 2012
Terminated
May 9, 2014
Recall Number
Z-1462-2012
Quantity
145
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

The monitor Ceiling suspension may fail, which may cause the monitor to drop.

Distribution

Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Algeria, Argentina, Belgium, Brazil, Canada, Czech Rep, Chili, China, Columbia, Denmark, Egypt, France, Germany, India, Indonesia, Italy, Japan, Malaysia, Mexico, Morocco, Netherlands, Oman, Russia, Saudi Arabia, Singapore, South Korea, Spain, Thailand, United Kingdom, and Vietnam.

Lot / Code Info

722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722001, 722003, 722008

Root Cause

Process control

Action Taken

Philips sent an "URGENT-FIELD SAFETY NOTICE" dated March 27, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter provides information for continued safe and proper use of the equipment. Contact your local Philips representative at 866-767-2822 for further support regarding this issue. Recall expanded 5-2013 to additional 460 units distributed worldwide.

More recalls by Philips Healthcare Inc.

View all recalls by this company →