SlideView VS200 slide viewer,

This recall is currently active, issued July 26, 2024. It was issued by Evident Scientific Inc.

Company
Evident Scientific Inc
Recall Initiated
July 22, 2024
Posted
July 26, 2024
Recall Number
Z-2416-2024
Firm Location
Waltham, MA
Official Source
View on FDA website ↗

Reason for Recall

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

Distribution

US

Lot / Code Info

65159 67105 67104 69153 67106 69154

Root Cause

Radiation Control for Health and Safety Act