Intellivue Information Center Model: M3150B#C01

This recall has been terminated (originally issued February 28, 2003).

Company
Philips Medical Systems, Inc. Cardiac & Monitoring Systems
Recall Initiated
January 2, 2003
Posted
February 28, 2003
Terminated
February 15, 2006
Recall Number
Z-0596-03
Quantity
39 units
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

Central Station Monitoring Information Centers may experience sreeen freeze, i.e. loss of display, keyboard or mouse lock up

Distribution

Nationwide Canada, Europe, Asia, Latin America

Lot / Code Info

Serial 4225A00106 4225A00107 4225A00108 4225A00111 4225A00112 4225A00113 4225A00114 4225A00116 4225A00117 4225A00118 4225A00119 4225A00127 4225A00145 4225A00149 4225A00150 4225A00151 4225A00152 4225A00164 4225A00165 4225A00166 4225A00167 4225A00169 4225A00188 4225A00189 4213A13698 4213A13700 4225A00122 4225A00136 4225A00138 4225A00137 4225A00139 4225A00140 4225A00142 4225A00143 4225A00144 4225A00101 4225A00102 4225A00103 4225A00104

Root Cause

Other

Action Taken

Philips Medical Systems notified users by Certified Mail on/about 2/19/03 that the firm was initiating a device field correction (recall) to conduct an upgrade to units witha HP VL400 or VL420 computer which is cconfigured for dual display by replacing the video and network communication card. Previous to this notification, Philips sent letters on 1/2/03 addressed to Biomedical Engineering Manager to advise users to reboot the system in the event of a system freeze.