RecallDepth

chromID Salmonella Agar, Reference No. 43621. A selective isolation and differentiation medium for the detection of Salmonella in human specimens.

This recall has been terminated (originally issued October 28, 2013).

Company
Biomerieux, Inc.
Recall Initiated
September 6, 2013
Posted
October 28, 2013
Terminated
June 24, 2014
Recall Number
Z-0087-2014
Quantity
90
Firm Location
Durham, NC
Official Source
View on FDA website ↗

Reason for Recall

Complaints for Salmonella strains growing on this medium producing an uncharacteristic white colored colony instead of the expected pale pink to mauve colored colonies. This could potentially lead to Salmonella colonies not being detected.

Distribution

Nationwide Distribution.

Lot / Code Info

Lot 1002234210 and 1002300040

Root Cause

Nonconforming Material/Component

Action Taken

Recall letters were issued to the US customers by Biomeriux, Durham, NC on 9/6/2013.

More recalls by Biomerieux, Inc.

View all recalls by this company →