Magic View 1000U, version VE40A. Digital Image Communication System

This recall has been terminated (originally issued July 19, 2003).

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
February 28, 2003
Posted
July 19, 2003
Terminated
August 25, 2004
Recall Number
Z-1029-03
Quantity
49 units
Firm Location
Malvern, PA
Official Source
View on FDA website ↗

Reason for Recall

software issue. New examinations may not be saved properly and the examination may be lost.

Distribution

The product was shipped to medical facilities in AL, CA, CO, FL, NE, IL, IN, MI, MN, NJ, NY, OH, PA, TN, and VA.

Lot / Code Info

Part number 7502003

Root Cause

Other

Action Taken

The recalling firm has issued a letter to their field service representative dated 2/28/02. The letter is entitled Safety Modification CN002/03/S which instructs the recalling firm's service engineers to visit each affected site and correct the software problem. A set of instructions were sent with the letter.