RecallDepth

Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0), STK-GT-001 (MT26403-1), STK-GT-008 (MT26403-5), STK-GT-013 (MT26403-1) , STK-GT-019 (MT26403-4) , STK-GT-100 (MT26403-2), STK-GT-109 (MT26403-2) , STK-GT-113 (MT26403-3)

This recall is currently active, issued June 16, 2025. It was issued by Dexcom Inc.

Company
Dexcom Inc
Recall Initiated
May 12, 2025
Posted
June 16, 2025
Recall Number
Z-1944-2025
Quantity
602,445 units
Firm Location
San Diego, CA
Official Source
View on FDA website ↗

Reason for Recall

Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

Distribution

worldwide distribution - US Nationwide and the countries of Andorra, Argentina Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, New Zealand, Norway, Oman, Poland, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.

Lot / Code Info

All Lots/ UDI: STK-AT-011 - 00386270002198 STK-AT-012 -00386270002518 STK-AT-013 - 00386270002525 STK-GT-001 - 00386270004291 STK-GT-008 - 00386270004260 STK-GT-013 - 00386270002273 STK-GT-019 - 04595986293080 STK-GT-100 - 00386270002020 STK-GT-109 - 00386270002129 STK-GT-113 - 00386270003638

Root Cause

Nonconforming Material/Component

Action Taken

On May 12, 2025, Dexcom issued a "Urgent: Medical Device Correction" Notification to affected consignees via E-Mail. On June 9, 2025, Dexcom issued an Urgent: Medical Device Correction update Dexcom asked consignees to take the following actions: 1. Verify if your receiver is affected by this issue: Check the serial number for any receiver(s) you use: https://dexcom.com/en-us/checkreceiver 2. If your receiver is affected, contact Dexcom Technical Support at 1-844-478-1600 to coordinate return of the affected receiver and a no charge replacement. 3. Whether or not your receiver is affected by this notification, Dexcom continues to recommend regular testing of the speaker. Test speaker function every time you charge your receiver. When you charge, a reminder prompt will invite you to test speaker function. 4.If you need assistance related to this issue or notification, please call Dexcom Technical Support at 1- 844-478-1600. Technical Support is available 24 hours a day/7 days a week to assist you.

More recalls by Dexcom Inc

View all recalls by this company →