Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250
- Company
- Bridge To Life Ltd
- Recall Initiated
- June 12, 2025
- Posted
- July 22, 2025
- Recall Number
- Z-2157-2025
- Quantity
- 4528 units
- Firm Location
- Northbrook, IL
Reason for Recall
The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024. Routine lot by lot testing was completed properly however the quarterly dose audit (sampling/testing to monitor the gamma sterilization process and confirm the validated parameters remain effective to assure sterility and bioburden reduction per requirements) was not completed on time.
Distribution
United States - New York, Maryland, Alabama, North Carolina, South Carolina, Ohio, Florida, Wisconsin, Utah, Tennessee, New Jersey, Georgia, Nebraska, Arizona, Minnesota, California, and Canada
Lot / Code Info
UDI 10860001991205-240206-260206-240372708 (case), 00860001991208-240206-260206-240372708 (bag), Lot Number 240372708; UDI 10860001991205-240212-260212-240402836 (case), 00860001991208-240212-260212-240402836 (bag), Lot Number 240402836; UDI 10860001991205-240214-260214-240402837 (case), 00860001991208-240214-260214-240402837 (bag), Lot Number 240402837.
Root Cause
Process control
Action Taken
Bridge to Life issued an IMPORTANT FIELD SAFETY NOTICE to its consignees on 06/12/2025 via email. The notice explained the issue and requested the following: Actions required by the customer: 1. To be aware and acknowledge this field safety notice. 2. We are requesting customers to conduct a review of the existing EasiSlush inventory to determine if you have any quantities of the following three lot numbers 240372708, 240372836, 240402837. 3. If you may have further distributed the product, we request that quantity be included in your review and please advise us on where your product is forwarded with contact information. 4. Immediately notify Bridge to Life of the quantities in inventory. Fill out FSN Customer Reply Form below and send it back to the email address Vigilance@B2LL.com 5. FSN Customer Reply Form must be returned no later than June 18, 2025. 6. Lastly, please confirm no adverse events related to the subject lots and advise if you have any additional questions or concerns. For questions, please contact Vigilance@B2LL.com or 800-667-9524.